MG Dew Point MonitorEasy Temperature and Humidity Monitoring

MG40 (For Atmospheric-Pressure Air Applications) /
MG40A-P (For Compressed Air Applications)

Humidity Display
0.1 to 99.9%, 1/10 Resolution (MG40)
0.1 to 50.0%, 1/10 Resolution (MG40A-P)
Dew Point Display
–40.0 to +40.0 ℃, 1/10 Resolution
  • –40.0 to –60.0 ℃ is for reference only.
Temperature Display
–20.0 to +80.0 ℃, 1/10 Resolution

This products is available for purchase online.

MG40 / MG40A-P
MG40 / MG40A-P

Specifications

MG40※4 MG40A-P※5
Main Unit Specifications Humidity Display: 3-digit LED display 0.1 to 99.9%, 1/10 resolution※1 0.1 to 50.0%, 1/10 resolution※1
Displays "––––" on out of range, short circuit, or disconnect.
Dew Point Display: 4-digit LED display
(1 digit for symbol)
–40.0 to +40.0 °C, 1/10 Resolution※1 ※2
(Displays "L" at –60 ℃ or lower, and "H" at +40 ℃ or higher)
Values of –40 ℃ and below are for reference only.
Temperature Display: 4-digit LED display
(1 digit for symbol)
–20.0 to +80.0 ℃, 1/10 Resolution
(Displays "L" at –20 ℃ or lower, and "H" at +80 ℃ or higher.
Displays "––––" on out of range, short circuit, or disconnect.)
Operable Ambient Temperature Range 5 to 40
Operable Ambient Temp Range %RH 0 to 85 (No condensation)
Storage Temp/Humidity Range ℃ % 5 to +55 / 45 to 85%
Power Supply V 100 Vac ±10%
Power Consumption W 20
External Dimensions Height mm 80
Depth mm 220
Width mm 260
Piping Connection Size mm Φ4 (One-touch fitting)
Mass kg 2.7 2.9
外部出力 Voltage Output Dew Point: 0 to 5 Vdc (–60 to +40 ℃)
Humidity: 0 to 5 Vdc (0 to 100%) Humidity: 0 to 2.5 Vdc (0 to 50%)
Temperature: 0 to 5 Vdc (–20 to +80 ℃)
Alarm Output Dew Point (Upper/Lower Limit) / Humidity (Upper/Lower Limit) / Temperature (Upper/Lower Limit)
No-voltage Contact Outputs (1a), 2 sets (AL1 and AL2)
Sensor Specifications Measurable Fluid Pure air (at atmospheric pressure) / ISO purity class 8 (free of water droplets, oil, suspended organic solvents, etc.)※6 Compressed air (free of water droplets, oil, dust. Always install a filter before this product.)※6
Pressure Range MPa Atmospheric Pressure 0.1 to 0.8
Temperature Gauge Precision ±1
Humidity Gauge Precision %RH ±2 (20 to 80%) at 25 ℃ ±2 (20 to 40%) at 25 ℃
Calculated Dew Point Precision ±3 for D.P. of –30 to +40 ℃ / ±5 for D.P. of –30 to –40 ℃※3
Values of –40 ℃ or lower are for reference only. (When the measured air temperature is 25 ℃.)
Sampling Flow Rate L/min 3 to 5 Using a fixed orifice purge.
  1. Pressing the display switch alternates display between humidity and dew point.
  2. The dew point is a calculated value based on temperature and humidity.
  3. The calculated dew point precision value at the time the product is shipped. This is not a guaranteed value.
  4. The MG40 is a dew point monitor intended for use with atmospheric pressure air. If taking a measurement, place the sensor directly in the air to be measured.
  5. The MG40A-P is a dew point monitor intended for use with compressed air. If taking measurements, then supply the air to be measured to the product via the included sampling air tube (Φ4, L2000).
  6. Using or storing in the following atmospheres will lead to deterioration of the sensor and necessitate recalibration in a short period of time:acetic acid, hydrochloric acid, ammonia, ethyl acetate, xylene, butanol, ethylene dichloride.