Thermal Stream Compact Thermostatic Chamber Constant Temperature and Humidity Chamber in a Compact and Space Saving Design

APS-200LL-D / APS-200LLP-E /
APS-200LLKP-E

Temperature Control Range
-40℃ to +100℃
Temperature Distribution
±0.5℃(-40℃ to +60℃)
±1.0℃(+61℃ to +80℃)
±1.5℃(+81℃ to +100℃)
Operable Temperature Range
5℃ to 35℃
  • The perfect machine to be used for every kind of testing and inspection on electrical component production lines, analysis inspection, temperature characterization, food-related storage confirmation, etc.
  • Internal operation is possible as the door is opened and closed under high and low temperature conditions.
  • Employing a motorized sliding front glass door, there is little temperature fluctuation when the door is opened and closed, and it's possible to view internal operations.
Thermal Stream Compact Thermostatic Chamber

Specifications

Model Constant Temperature Model Constant Temperature, Constant Humidity Model
APS 200LL-D 200LLP-E 200LLKP-E
Instrumentation Settable Digital Temperature Controller Touch Panel with Programmable Controller
Performance Specifications
※1
Temperature Control Range※2 -40℃ to +100℃
Humidity Control Range 20% to 95% (When temp is 50 °C to 90 °C)
Temperature Distribution※3 ±0.5 °C (when temp is –40 °C to +60 °C),
±1.0 °C (when temp is +61 °C to +80 °C),
±1.5 °C (when temp is +81 °C to +100 °C)
Humidity Distribution※3 ±5%
Temperature Rise Time※4 60 min or less (–40 °C → +100 °C)
Temperature Fall Time※4 60 min or less (+20 °C → –40 °C)
Control Method PID Control / Pt100Ω / Resolution: 0.1 °C
Operable Temperature Range 5℃ to 35℃
Electrical Specifications Power Supply Three-phase 200V±10%、50/60Hz
Maximum Operating Current 23A 26.5A
Power Supply Capacity 8kVA 9.2kVA
Circuit breaker Built in
Refrigerant R404A
Chiller Compressor Output
(2 built-in compressors)
No. 1: 1.1 kW, No. 2: 1.1 kW
Internal Construction SUS304
Mass 300kg 320kg
Touch Panel Programmable Programmable: 20 programs (program linking possible) /
Step count: 20 steps (per program) / Program repeat count: 9999 times /
Step time setting: 0 to 9999 min (minimum 1 min) /
Other functionality: wait function, hold function, skip function
External Dimensions (W×D×H)※5 970mm×1010mm×1625mm
  1. Capacity is when doors are fully closed.
  2. Temperature range is when under no load, no test sample present, operating at the rated voltage, temperature is where air is being blown out inside the chamber, at an ambient temperature of +10 °C to +25 °C.
  3. The temperature and humidity divergence values are based on the JTM standard of an ambient temperature of +10 °C to +25 °C, when under no load, when there is no test sample present, under stable conditions. (The value is the average of the absolute value of the averages of the positive and negative values of 8 corner points and the center of the sample chamber.)
  4. The temperature rise and fall times are when the ambient temperature is +20 °C, with no load, when there is no test sample present, at the rated voltage, tested at the air outlet point inside the chamber when operating under maximum capacity.
  5. Does not include protruding parts.
  • There may be a 2 °C to 7 °C difference between the temperature inside the chamber and the temperature of air being blown into chamber due to intake of external heat.